Test Code AL CONUT Allergen Coconut IgE
Clinical System Name
Allergen Coconut IgE
Sample Requirements
Specimen:Whole Blood
Container(s):Gold Top SST, Red
Preferred Vol:2 mL
Minimum Vol:1 mL
Note:
Processing Instructions
Reject due to:
Spin:Y
Aliquot:Y
Temp:2-8C
Storage location:CPA Refrigerator, FrigC rack, with requisition
Off-site collection:Spin blood and refrigerate aliquot with requisition
Note:Each additional allergen requires 50mcL in addition to the inital 0.3mL. 10 allergens would require 0.8 mL minimum volume.
Stability
| Specimen Type | Temperature | Time | 
|---|---|---|
| Room temp | ||
| Refrigerated | 7 d | |
| Frozen | 6 m | 
Performing Laboratory
Seattle Children's Laboratory
Department
Department: Chemistry
Location: Chem East
Phone Number: (206) 987-3694
Synonyms
RAST Coconut IgE
Coconut IgE Allergen
Availability
| STAT | Performed | TAT | 
|---|---|---|
| N | M, W, F | 2-4 d | 
Methodology
Method: 3rd generation Chemiluminescent Immunoassay on Siemens Immulite 2000 XPI
Analytical Volume: 0.3 mL
Limitations:
Reference Range
| 
 Class  | 
 IgE kU/L  | 
 Allergen Reactivity  | 
| 
 0  | 
 <0.35  | 
 Negative  | 
| 
 1  | 
 0.35-0.69  | 
 Low  | 
| 
 2  | 
 0.70-3.49  | 
 Moderate  | 
| 
 3  | 
 3.50-17.49  | 
 High  | 
| 
 4  | 
 17.50-52.49  | 
 Very High  | 
| 
 5  | 
 52.5-99.9  | 
 Very High  | 
| 
 6  | 
 >100  | 
 Very High  | 
CPT Codes
86003