Sign in →

Test Code LAB1050 Manganese, Serum

Additional Codes

MANGANESE

Reporting Name

Manganese, S
Seattle Children's Hospital Note:

SCH Clinical System Name: MANGANESE

Useful For

Monitoring manganese exposure using serum specimens

 

Nutritional monitoring


Specimen Required


Patient Preparation: High concentrations of gadolinium, iodine, and barium are known to interfere with most metal tests. If gadolinium-, iodine, or barium-containing contrast media has been administered, a specimen should not be collected for 96 hours.

Supplies: Metal Free Specimen Vial (T173)

Collection Container/Tube: Plain, royal blue-top Vacutainer plastic trace element blood collection tube

Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial

Specimen Volume: 0.3 mL

Collection Instructions:

1. Allow the specimen to clot for 30 minutes, and then centrifuge to separate serum from the cellular fraction. Serum must be removed from cellular fraction within 4 hours of collection. Avoid hemolysis.

2. Remove the stopper. Carefully pour specimen into a Mayo metal-free, polypropylene vial, while avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.

3. See Metals Analysis Specimen Collection and Transport for complete instructions.


Seattle Children's Hospital Note:

Collect 0.6 mL whole blood in a Royal Blue top trace element tube.

 

Processing Note: Use of a standard pipette is acceptable for aliquoting these samples.

Specimen Type

Serum

Specimen Minimum Volume

0.2 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 28 days METAL FREE
  Ambient  28 days METAL FREE
  Frozen  28 days METAL FREE

Day(s) Performed

Tuesday

Seattle Children's Hospital Note:

TAT: 3 - 8 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Method Name

Triple-Quadrupole Inductively Coupled Plasma-Mass Spectrometry (ICP-MS/MS)

Method Description

The metal of interest is analyzed by triple-quadrupole inductively coupled plasma mass spectrometry.(Unpublished Mayo method)

Reference Values

0-17 years: Not established

≥18 years: 0.5-1.2 ng/mL

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

83785

LOINC Code Information

Test ID Test Order Name Order LOINC Value
MNS Manganese, S 5683-8

 

Result ID Test Result Name Result LOINC Value
8413 Manganese, S 5683-8

Report Available

2 to 8 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus OK
Seattle Children's Hospital Additional Information:

Aliases: Manganese (Mn); Mn (Manganese)