Sign in →

Test Code LAB1976 Urine Drug Abuse Screen-Rapid (1 hr)

Clinical System Name

Urine Drug Abuse Screen-Limited (1 hr)

Synonyms

LAB1976

DRUG ABUSE

Rapid Urine Drug Abuse Screen

Drug Abuse Standard Screen, Urine

Drug Abuse Screen, Urine
Drug Screen, Standard Urine
PCP Screen, Urine
THC Screen, Urine
Tox Screen, Urine - Drug Abuse Standard
Urine Drug Abuse Screen
Urine Tox Screen

Sample Requirements

Specimen: Urine

Container(s): Sterile Screw-Capped Container, Sterile Screw-Capped Tube, Syringe

Preferred Vol: 2 mL

Minimum Vol: 0.5 mL

 

Note: This test provides a preliminary analytical result that should not be used for legal purposes. A confirmed analytical result can be obtained by more specific testing such as gas chromatography/mass spectrometry at an additional charge. Samples needing confirmation will be sent to Harborview. Clinical judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. A negative result does not exclude the presence of a particular drug due to different sensitivity of each drug. Contact SCH Lab to arrange for confirmation test. If urine is frozen, it may only be thawed once prior to testing.

Processing Instructions

Reject due to: Non-sterile container, age of Specimen, Insufficient quantity

Spin: N

Aliquot: N

Storage location: Core 5 Chemistry Refrigerator

 

Off-site collection: Sample should be refriegerated and sent to main hospital lab.

Stability

Specimen Type Temperature Time
Urine RT

≤ 2 h

Urine 2-8 C

≤ 2 d

Urine -20 C or -70 C

≤ 30 d

*** Stability times come from MEDTOXScan. If urine is frozen, it may only be thawed once prior to testing.

Availability

STAT Performed TAT
Y 24/7 1 h

Performing Laboratory

Seattle Children's Laboratory    

Department

Department:  Core Chemistry

Phone Number206-987-2617 (Client Services)

CPT Codes

80306

Methodology

Method: One step immunochromatographic assay with MEDTOXScan test device and MEDTOXScan reader.

Analytical Volume: 0.5 mL

 

Method Specifications:  https://www.medtoxdiagnostics.com/wp-content/uploads/102038-Package-Insert-PV-Test-System.pdf

 

Reference Range

NEGATIVE

 

Detection is based on the following drugs at the following cut-off levels.

 

Amphetamines (AMP)                       d-Amphetamine                         500 ng/mL
Barbiturates (BAR) Butalbital 200 ng/mL
Benzodiazepine (BZO) Nordiazepam 150 ng/mL
Buprenorphine (BUP) Buprenorphine 10 ng/mL
Cocaine (COC) Benzoylecgonine 150 ng/mL
Methamphetamine (MAMP) d-Methamphetamine 500 ng/mL
Methadone (MTD) Methadone 200 ng/mL
Opiates (OPI) Morphine 100 ng/mL
Oxycodone (OXY) Oxycodone 100 ng/mL
Phencyclidine (PCP) Phencyclidine 25 ng/mL
THC Cannabinoids (THC) 11-nor-9-carboxy-∆9-THC 50 ng/mL
Tricyclic Antidepressants (TCA) Desipramine 300 ng/mL

 

Description

This test provides a preliminary analytical result that should not be used for legal purposes. Qualitative screening for amphetamines, methamphetamine, barbiturates, benzodiazepines, cannabinoids, cocaine, opiates, phencyclidine, methadone, propoxyphene, buprenorphine, Oxycodone, and tricyclic antidepressants.