Test Code LAB210 Body Fluid, Cell Count
Clinical System Name
Body Fluid Cell Count
Synonyms
Body Fluid, Cell Count
BF CELLCT
LAB210
Sample Requirements
Specimen: Body Fluid
Container(s): Sterile Screw-Capped Container/Tube or Syringe, Lavender/EDTA Tube*, Dark Green/Sodium Heparin tube*
Preferred Vol: 2.0 mL
Minimum Vol: 0.5 mL
Notes: Note type of body fluid on requisition and container. Send to the lab immediately for accurate results, cells start to lyse upon standing.
For bloody fluids or synovial fluid CELL COUNTS, collect in a lavender EDTA tube to prevent clotting.
Collect synovial fluid for crystal exams in dark green Sodium Heparin tube. See Crystals, Fluid. Lavender/EDTA and Lithium Heparin tubes may produce crystalline structures resembling monosodium urate in synovial fluid and are therefore not recommended for crystal analysis.
Processing Instructions
Deliver to Core lab Hematology bench ASAP.
Reject due to: Absess or wound source; Age of Specimen (give to Core Lab for determination); Clotted Solid
Spin: N
Aliquot: N
Storage location: Core 5 Fridge
Off-site collection: Sample should be sent to main campus lab as soon as possible (direct).
Stability
Specimen Type | Temperature | Time |
---|---|---|
Body Fluid | RT |
≤ 1 h |
Body Fluid | 2-8 C |
≤ 1 h |
-20 C or -70 C |
N |
Availability
STAT | Performed | TAT |
---|---|---|
Y | 24/7 | 1 h |
Performing Laboratory
Seattle Children's Laboratory
Department
Department: Core Hematology
Phone Number: 206-987-2617 (Client Services)
CPT Codes
89050: Cell Count, NO Differential
89051: Cell Count, WITH Differential
88108: Cytology
Methodology
Method: Manually counted via plating on an iNCYTO hemocytometer and/or Sysmex XN-3000: RF/DC detection method, Flow cytometry with semiconductive laser, photometrically, SLS-hemoglobin method, hydrodynamic focusing
Analytical Volume: 0.05mL
Reference Range
Reference ranges are not available for Body Fluid.