Sign in →

Test Code LAB2589 % Oxygen Saturation Venous PRE oxygenator

Important Note

Avoid air bubbles.

Place specimen in outside pocket of a pre-filled biohazard bag of ice, then put into a separate biohazard bag for transport.

Clinical System Name

% Oxygen Saturation Venous PRE oxygenator

Synonyms

LAB2589

O2 Sat V PRE

O2 Sat Venous PRE oxygenator

PRE oxygenator O2 Sat Venous

Sample Requirements

Specimen:  Whole Blood

Container(s):  Heparinized Syringe

Preferred Vol: 1.0 mL venous blood in a Heparinized Syringe

Minimum Vol: 0.5 mL venous blood in a Heparinized Syringe  

 

Note:  Expel air from syringe and cap it immediately after obtaining the sample. Place specimen in OUTSIDE pocket of a pre-filled biohazard bag of ice, then put into a separate biohazard bag for transport.

Processing Instructions

Deliver to Core Lab on ice. Notify technologist - STAT

 

Reject due to:  Inadequately filled specimens, air bubbles, visible clots, age of specimen

Spin:  N

Aliquot:  N

Temp: 0-4 C

Stability

Specimen Type Temperature Time
Heparinized Syringe RT ≤ 15 min
Heparinized Syringe 0-4 C ≤ 30 min
  -20 C or -70 C N

Availability

STAT Performed TAT
Y 24/7 15 min

Performing Laboratory

Seattle Children's Laboratory    

Department

Department:  Chemistry - Core

Phone Number: 206-987-2617 (Client Services)

CPT Codes

82810

Methodology

Method:  Oximetry by SIEMENS RAPIDPOINT 500

 

Analytical Volume:  0.5 mL Heparinized Syringe

Reference Range

Age % Saturation
0 - 1 d  85 - 90
> 1 d

95 - 98

 