Test Code LAB3553 Warfarin Prothrombin Time + INR

Clinical System Name

Warfarin Prothrombin Time + INR
 

Synonyms

LAB3553

WARF PT

INR
Protime
PT
PT INR

Sample Requirements

Specimen: Whole Blood

Container(s): Lt. Blue/Citrate

Required Vol: one 1.8 mL tube or one 2.7 mL tube (do NOT over or under fill)

Minimum Vol: N

 

Note: This is enough for a WARF PT, PTT, TT, and Fibrinogen on the same sample. Preferred method of collection is venipuncture with vacuum fill. Test results are affected by incorrect blood volume. Use of a Vascular Access Device for the collection of coag testing is not recommended.

If this method is used it requires a 5 cc clearing volume; 3 cc for size 2 French catheter or smaller. Blood must be transferred to the Lt Blue/Citrate tube by use of a blood transfer device to ensure proper fill. 

Specimen should be centrifuged within 1 hour of collection.

 

Processing Instructions

Main Campus CPA: Core Technologist will process specimen. Deliver blood directly to Core Coag bench.

 

Reject due to: Clotted, Insufficient quantity (underfill), or Improper collection (overfill).

Spin: Y

Aliquot: Y

Storage location: Core 14 Freezer ( -70 C).

 

Core Technologist: Specimen should be centrifuged within 1 hour of collection. Spin whole blood, remove plasma. Transfer upper 3/4 layer of plasma to plastic tube affixed with large computer label.

 

Off-site collection: Preferred method of collection is venipuncture with vacuum fill. Test results are affected by incorrect blood volume, note the fill line on the tube. Use of a Vascular Access Device for the collection of coag testing is not recommended. If this method is used it requires a 5 mL waste volume; 3 mL for size 2 French catheter or smaller. Blood must be transferred to the Lt Blue/Citrate tube by use of the blood transfer device.

 

Specimen should be centrifuged within 1 hour of collection. Double spin, transfer upper 3/4 layer of plasma to plastic tube affixed with large computer label. Freeze until solid one aliquot of 1.0 mL plasma in plastic tube at -70C (preferred) or -20C.

 

Non-Children's Hospital Off-site collection: Ship completely frozen on dry ice.

 

Children's Regional Clinics: Ship completely frozen in frozen Nalgene Labtop cooler with ice pack inside an insulated soft cooler.

Stability

Specimen Type Temperature Time
Citrated platelet-poor plasma Room temp ≤4 h
  Refrigerated N
  Frozen   -20 C or -70 C ≤3 m

 

Availability

STAT Performed TAT
Y

Daily

1 h

 

Performing Laboratory

Seattle Children's Laboratory    

 

Department

Department:  Coagulation

Phone Number: 206 987-2617 (Client Services)

CPT Codes

85610

Methodology

Method: Electromagnetic mechanical clot detection system by STA-R MAX

Analytical Volume: 1 mL plasma

This is enough for a WARF PT, PTT, TT, and Fibrinogen on the same sample.

 

Limitations: Method is affected by improper collection, processing, and storage.

High hematocrit greater than or equal to 56% requires a citrate adjusted tube.

Reference Range

Reference range determined by provider.

 

Description

The Warfarin PT test is used to monitor patients receiving Warfarin therapy.

Critical Values

 

Critical INR Critial PT
Less than or equal to 1.5 or greater than or equal to 4.0 >39.2 s